Residual Risks

Electric leakage causes electrical shock.

The product uses a battery power supply that will not cause electrical shock due to leakage.

Acceptable

Incorrect cuff pressure results in incorrect blood pressure values.

The user manual states that if you have doubts about blood pressure, you should inform your doctor.

Acceptable

Unable to automatically relieve pressure, causing the user discomfort.

The user manual states that if you feel uncomfortable, you should stop the measurement.

Acceptable

The product is damaged during the delivery process and fails.

If the product is damaged due to shipping, please contact the distributor.

Acceptable

Causes adverse skin reactions.

All materials that come in contact with human skin have passed the biocompatibility test.

Acceptable

Electromagnetic fields interfere with other product functions.

The products have passed EMC EN 60601-1-2 and IEC 301489-1/-17 verification tests.

Acceptable

The product function is affected by the interference of environmental electromagnetic fields.

The products have passed EMC EN 60601-1-2 and IEC 301489-1/-17 verification tests.

Acceptable

Generates an electromagnetic interference that affects other nearby products.

The products have passed EMC EN 60601-1-2 and IEC 301489-1/-17 verification tests.

Acceptable

Excessive pressure causes bruising or redness.

The product has an over-pressure protection mechanism. If you feel unwell during the measurement, please remove the cuff and stop using the device.

Acceptable

Unable to automatically release pressure, causing the user discomfort.

The product has an over-pressure protection mechanism. If you feel unwell during the measurement, please remove the cuff and stop using the device.

Acceptable

Incorrect use that result in incorrect blood pressure measurement.

The user manual explains the correct way to use the device.

Acceptable

The blood pressure value measured (on OLED) by the device displays an incorrect error value, resulting in false judgement.

The product has passed the wireless interference test and meets the requirements of EN 300 328 regulations.

Acceptable

Causes unstable blood pressure.

The user manual explains error messages and precautions.

Acceptable

May cause injury or incorrect blood pressure.

The user manual explains error messages and precautions.

Acceptable